▼ DC Calibration
The device manual contains a detailed description of the DC calibration!
- Select the DC analog output to be calibrated.
- The "Set" field displays either "-20 mA" or "-10 V". This value cannot be changed.
Note: If the output range is set to 0 mA to 20 mA, 4 mA to 20 mA or -20 mA to 20 mA, the range from -20 mA to 20 mA is calibrated. If the output range is set to 0 V to 10 V or -10 V to 10 V, the range from -10 V to 10 V is calibrated. |
- Enter the maintenance password into the "Password" field and click the "Calibrate" button.
- Enter the measured value displayed by the measuring device into the "Measured" field.
- Enter the maintenance password into the "Password" field and click the "Calibrate" button.
- The "Set" field displays either "20 mA" or "10 V". This value cannot be changed.
- Enter the maintenance password into the "Password" field and click the "Calibrate" button.
- Enter the measured value displayed by the measuring device into the "Measured" field.
- Enter the maintenance password into the "Password" field and click the "Calibrate" button.
This action is protected. Enter the correct password.
Password
Start: Press "Calibrate" button to start the calibration.
Output | |
Last calibration dates
|
-20 mA to 20 mA | -10 V to 10 V |
K2/3 | 2010-04-09 15:21:21:631 |
2010-04-09 15:15:40:676 |
K4/5 | 2010-04-09 15:20:34:312 |
2010-04-09 15:16:26:967 |
K6/7 | 2010-04-09 15:19:53:267 |
2010-04-09 15:17:10:873 |
K8/9 | 2010-04-09 15:19:07:342 |
2010-04-09 15:17:51:622 |