▼ DC Calibration

The device manual contains a detailed description of the DC calibration!

  1. Select the DC analog output to be calibrated.
  2. The "Set" field displays either "-20 mA" or "-10 V". This value cannot be changed.
    Note: If the output range is set to 0 mA to 20 mA, 4 mA to 20 mA or -20 mA to
    20 mA, the range from -20 mA to 20 mA is calibrated. If the output range is set
    to 0 V to 10 V or -10 V to 10 V, the range from -10 V to 10 V is calibrated.
  3. Enter the maintenance password into the "Password" field and click the "Calibrate" button.
  4. Enter the measured value displayed by the measuring device into the "Measured" field.
  5. Enter the maintenance password into the "Password" field and click the "Calibrate" button.
  6. The "Set" field displays either "20 mA" or "10 V". This value cannot be changed.
  7. Enter the maintenance password into the "Password" field and click the "Calibrate" button.
  8. Enter the measured value displayed by the measuring device into the "Measured" field.
  9. Enter the maintenance password into the "Password" field and click the "Calibrate" button.

This action is protected. Enter the correct password.

Password


Start: Press "Calibrate" button to start the calibration.

Output
Set <--
Measured


Last calibration dates

-20 mA to 20 mA-10 V to 10 V
K2/3 2010-04-09 15:21:21:631 2010-04-09 15:15:40:676
K4/5 2010-04-09 15:20:34:312 2010-04-09 15:16:26:967
K6/7 2010-04-09 15:19:53:267 2010-04-09 15:17:10:873
K8/9 2010-04-09 15:19:07:342 2010-04-09 15:17:51:622